740 Portable Suspended Solids
Note: The factory default password is 1000, if the user password has been
forgotten please contact Partech technical support.
9.6 Monitor Adv Config
9.6.1 Factory Cal Values (1 to 5)
These values are used to store the factory calibration data of the attached sensor. The
only time that these values will need to be accessed is when a replacement sensor is
fitted to the monitor. The new sensor will be supplied with a set of calibration values
that must be entered - failure to do so will affect the calibration and measurement
accuracy of the instrument.
9.7 Load Monitor Defaults
It is sometimes necessary to restore the default values for all the user configurable
parameters such that the instrument is in a known state. To load the monitor defaults,
select Load Monitor Defaults and press OK. To prevent accidental use, the user
password must be entered prior to confirming the operation.
Page 24 of 46 172430IM-Iss07 Issue Date 23/10/2014