Basic Installation
25
Unpacking and Inspection
1. Unpack the components carefully.
Keep the packing materials for possible future storage or reshipment.
2. Examine the components for any signs of damage in shipment.
In the event of damage or missing parts, file an immediate claim with the
authorized carrier, and inform your PerkinElmer office or representative.
After the instrument has been unpacked, check the exterior and interior for possible
damage as follows:
1. Check the entire outer cabinet of the spectrometer for damage.
2. Make sure that terminals, fuse holders, etc. are not damaged.
3. Open the sample compartment cover, checking that it moves freely without
binding.
The compartment must be free of dust or other foreign matter.
4. Close the sample compartment cover.
5. Open the lamp compartment as described in Halogen Lamp Replacement on
page 56.
6. Check for damages or loose cables.
7. Close the lamp compartment.
NOTE: We recommend that the initial installation of the spectrometer should be performed
by a PerkinElmer service representative. After receipt of the instrument, please
contact your PerkinElmer office or representative for further information.
You are required to prepare a suitable working space.