Lambda 20, Lambda 40 UV/Vis Spectrometers
12
Environment
Operating Conditions
WARNING
Explosive Atmosphere
This instrument is not designed for operation in an explosive
atmosphere.
The instrument will operate correctly under the following conditions:
• Indoors.
• Ambient temperature +15 ºC to +35 ºC.
• Ambient relative humidity 20% to 80%, without condensation.
• Altitude in the range 0 m to 2000 m.
Storage Conditions
You can store the instrument safely under the following conditions:
• Indoors.
• Ambient temperature +15 ºC to +35 ºC.
• Ambient relative humidity 20% to 80%, without condensation.
• Altitude in the range 0 m to 2000 m.
When you remove the instrument from storage, before putting it into operation allow
it to stand for at least a day under the approved operating conditions.