Maintenance
53
2. Using a soft cloth and mild laboratory detergent solution, lightly scrub away all
foreign material.
3. Using a clean cloth dampened with water, rinse the cleaned surfaces
thoroughly.
4. Dry with lint free cloth or tissue.
Sample Compartment Window
Windows are provided with the spectrometer. The window is made of silica and may
be used in the entire spectral range of the spectrometer.
The window seals the sample compartment and thus protects the instrument’s optics
from dust and fuming or aggressive samples.
Generally, the window should be installed at all times.
The window is an optical component and requires the same care and handling
as cells.
You can remove the window to clean it. The frame is magnetic and can be
removed by hand.
Windows are most suitably cleaned by wiping them with a soft cloth moistened
with ethanol.