Maintenance
61
4. Unplug the deuterium lamp connector from the terminal board by squeezing in
the two lugs at each side of the connector and carefully pulling the connector
vertically upward.
5. Remove the lamp assembly from the bracket by slackening the thumbscrew
and pulling the lamp mount vertically upward.
Save the thumbscrew for use with the new lamp assembly.
Lamp baffle
Thumbscrew
(deuterium lamp)
Deuterium lamp
connector
Deuterium lamp
assembly