EasyManuals Logo

Perkin-Elmer Lambda 20 User Manual

Default Icon
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Lambda 20, Lambda 40 UV/Vis Spectrometers
84
Optics
Beam center height 15 mm above cell holder bottom
Beam cross-section
1 mm slit ca. 0.6 mm x 9 mm (width x height)
2 mm slit ca. 1 mm x 9 mm (width x height), at
focal point of sample and reference
beam in sample compartment
Optical pathlength in
sample compartment
121 mm
Grating (Monochromator)
Holographic concave grating with 1053 lines/mm
in the center
Radiation sources Prealigned deuterium and halogen lamps
Detector
Photodiodes
(One for the sample beam and one for the
reference beam)
Abscissa
Wavelength range
190 n to 1100 nm;
0 nm for alignment purposes
Wavelength accuracy ±0.3 nm
Wavelength reproducibility ±0.1 nm
Spectral bandwidth 1 nm or 2 nm (fixed slit)
Lamp change
Automatically at 326 nm
(selectable over the whole wavelength
range, see Super User Factory
Configuration)
Scan speeds
7.5, 15, 30 , 60, 120, 240, 480, 960,
1920 and 2880 nm/min
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Perkin-Elmer Lambda 20 and is the answer not in the manual?

Perkin-Elmer Lambda 20 Specifications

General IconGeneral
BrandPerkin-Elmer
ModelLambda 20
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals