Lambda 20, Lambda 40 UV/Vis Spectrometers
84
Optics
Beam center height 15 mm above cell holder bottom
Beam cross-section
1 mm slit ca. 0.6 mm x 9 mm (width x height)
2 mm slit ca. 1 mm x 9 mm (width x height), at
focal point of sample and reference
beam in sample compartment
Optical pathlength in
sample compartment
121 mm
Grating (Monochromator)
Holographic concave grating with 1053 lines/mm
in the center
Radiation sources Prealigned deuterium and halogen lamps
Detector
Photodiodes
(One for the sample beam and one for the
reference beam)
Abscissa
Wavelength range
190 n to 1100 nm;
0 nm for alignment purposes
Wavelength accuracy ±0.3 nm
Wavelength reproducibility ±0.1 nm
Spectral bandwidth 1 nm or 2 nm (fixed slit)
Lamp change
Automatically at 326 nm
(selectable over the whole wavelength
range, see Super User Factory
Configuration)
Scan speeds
7.5, 15, 30 , 60, 120, 240, 480, 960,
1920 and 2880 nm/min