EasyManuals Logo

PerkinElmer EnVision User Manual

PerkinElmer EnVision
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
Specifications
74
TRF
DELFIA
100 55 s 2 m 32 s 8 m 52 s
AlphaScreen
Std
- 1 m 32s 4 m 54 s 19 m 28 s
AlphaScreen
HTS
- 51 s 1 m 52 s 9 m 23 s
Measurement Performance
Technology Feature Performance
Fluorescence
intensity, filters
Detection limit (96-
plate, 200 µL)
< 4 pM, <0.8 fmol/well,
fluorescein, 100 flashes from top
Fluorescence
intensity, filters
Detection limit
(384-plate, 50 µL)
< 4 pM, <0.2 fmol/well,
fluorescein, 100 flashes from top
Fluorescence
intensity, filters
Detection limit
(1536-plate, 7.5
µL)
< 20 pM, <0.15 fmol/well,
Fluorescein, 100 flashes from top
Fluorescence
intensity,
monochromator
Detection limit (96-
plate, 200 µL)
< 10 pM, <2 fmol/well,
Fluorescein, 100 flashes from top
Fluorescence
intensity,
monochromator
Detection limit
(384-plate, 50 µL)
< 80 pM, <4 fmol/well,
Fluorescein, 100 flashes from top
Fluorescence
intensity,
monochromator
Bandwidth
(Excitation /
Emission)
< 8nm
Fluorescence
polarization
Detection limit (96-
plate, 200 µL)
< 3 mP, 1nM Fluorescein, 100
flashes
Fluorescence
polarization
Precision (384-
plate, 50 µL)
< 3 mP, 1nM Fluorescein, 100
flashes
Fluorescence
polarization
Precision (1536-
plate, 7.5 µL)
< 7 mP, 1nM Fluorescein, 100
flashes
Time-resolved
fluorescence
Detection limit (96-
plate, 200 µL)
< 55 fM, Eu, <11 amol/well, 100
flashes from top
Time-resolved
fluorescence
Detection limit
(384-plate, 50 µL)
< 55 fM, Eu, < 2.8 amol/well, 100
flashes from top

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer EnVision and is the answer not in the manual?

PerkinElmer EnVision Specifications

General IconGeneral
Light SourceXenon flash lamp
SoftwareEnVision Manager
Read ModesAbsorbance, Fluorescence, Luminescence, Time-Resolved Fluorescence, Fluorescence Polarization
Plate Formats384-well
Detection TechnologyPMT
Temperature ControlAmbient to 45°C