Installing your Lambda Spectrometer . 51
NIR Wavelength:
Select Manual set, Manual set 1 peak, Manual set 2 peaks,
Auto search 1 peak, or Auto search 2 peaks.
Manual set enables you to enter the offset and factor (without the need to
determine the values in advance).
Manual set 1 peak enables you to change the offset. A well known peak can be
shifted to an exact wavelength by entering values of Old Peak 1 and
New Peak 1.
Manual set 2 peaks enables you to change the offset and factor by specifying
the old and new values of 2 well known peaks by entering values of
Old Peak 1, New Peak 1, Old Peak 2 and New Peak 2.
Auto search 1 peak enables you to change the offset. The software performs an
automatic search for the D
2
peak at 1312.2 nm. The measured peak is then
shifted to the exact wavelength.
Auto search 2 peaks enables you to change the offset and factor. The software
performs an automatic search for the D
2
peak at 1312.2 nm and the peak at
0.0 nm. The measured peaks are then shifted to the exact wavelengths.
NOTE: Auto search 2 peaks is the default setting.
Slits:
There are no user-defined settings in the Slits calibration.
Zero Percent T:
There are no user-defined settings in the Zero Percent T calibration.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda Series and is the answer not in the manual?

PerkinElmer Lambda Series Specifications

General IconGeneral
CategoryUV Vis Spectrophotometer
Spectral Bandwidth0.5, 1, 2, 4 nm
Data Interval1, 2, 5 nm
Slit Width0.5, 1, 2, 4 nm (selectable)
Light SourceDeuterium and Tungsten Halogen lamps
DisplayLCD
InterfaceUSB, RS-232
Power Requirements100 - 240 VAC, 50/60 Hz
Wavelength Range190-1100 nm

Related product manuals