Maintenance
89
10.12 Maintenance and service schedule
Assembly
Part number
Quantity
per
unit
Function Operating time in
hours/weeks
Operating
Manual
Rep.
level
Place
Test facilities
Comments
Material
Designation
Part No.
160 8600 17200 <25800
1 52 104 <208
GSD 320 Basic unit, housing
Ventilator
51-0004
1 Clean filter mat X(3) I os Depending on severity
of contamination
Replace filter mat X I os Concurrent with
overhaul
Set of filter mats (10 pcs)(V) BN 846 136 -T
GSD 320 Basic unit, vacuum system
MVP 020/3 1 Replace diaphragm X(1) PK 0217 BN II os Usual lifetime of
>10000 operating hours
Wearing parts kit PU E22 003-T
Replace pump II os In case of malfunction Exchange MVP 020 PK T01 151 #A
HiPace 80 1
Change operating fluid X(3) PT 0208 BN II os Special tool, at least
every 4 years
Lubricant reservoir PM 143 740-TT
Replace bearing X(3) II ws At least every 4 years
or after 3 years of idle
time
Exchange pump PM P02 812C#A
Exchange pump II os In case of malfunction Exchange HiPace 80 PM P02 812 C#A
PKR 251
PT R26 002
1 Cleaning, alignment X BG 805 119 BN II os Concurrent with
overhaul
Maintenance kit BN 846 239 -T
Analyzer 1 Replace Filament BG 5214 II os In case of malfunction Tungsten filament
Yttriated Iridium filament
BN 846 281-T
BN 846 395-T
GSD 320 Gas inlet
Of ThermoStar 1 Replace orifice and capillary X(3) I os or in case of
malfunction
Spare parts kit PT 162 002
Replace gas inlet II os In case of malfunction Replacement gas inlet PT 162 003
Of OmniStar 1
Replace plunger and seals
Replace orifice
X(3) II os Or in case of
malfunction
Spare parts kit PT 162 001
Replace capillary X(3) I os Or in case of
malfunction
Spare parts kit PT 162 001
Replace gas inlet II os In case of malfunction Replacement gas inlet PT 162 004
GSD 320 Accessories
Calibration Device 1 Replenish II os When necessary 5 ml PFTBA B 2705 421
Replace glass tube II os In case of malfunction From spare parts kit BN 841 310 -T
Replace plunger and seals X(3) II os In case of malfunction Seals Kit BN 841 310 -T