3 Product description
3.1 Identifying the product
►
To ensure clear identification of the product when communicating with Pfeiffer Vacuum, always
keep all of the information on the rating plate to hand.
►
Learn about certifications through test seals on the product or at www.certipedia.com with compa-
ny ID no. 000021320.
3.2 Product features
Feature TCP 350
Connection panel Standard
suitable for Turbopumps from Compact series
Turbopumps from HiPace series
Mains frequency 50/60 Hz
Mains connection voltage 95 – 265 V AC
Current consumption, max. 4 A
Power consumption, max. 420 VA
Internal fuse 10 A, fast-acting
Tbl. 4: Product features
3.3 Function
The electronic drive unit TCP 350 is used for external voltage supply, control and monitoring of turbo-
pumps and connected peripheral devices. It is also possible to connect a pressure gauge.
Mod: TCP 350
Mod.-Nr.: P M C01 740
Inp ut: 1 15/ 230V -20 /+1 5% 420 VA
Outpu t: 0 - 72 V 6,5 VA
0-1500 Hz
Made in Germa ny 2014 /08
Ser. Nr.:78514xxx
REMOTE
SERVICE
RS 485
PUMP
GAUGE
AC INPUT
POWER
TCP 350
1
2
3
6
9
4
10
11
12
13
8
5
7
Fig. 2: TCP 350 connections and manual control elements
1 LC display, illuminated 8 "RS 485" connection
2 Status symbols 9 "SERVICE" connection
3 LED operating mode display 10 "PUMP" connection
4 Controls 11 "GAUGE" connection
5 Grounding connection PE 12 "AC INPUT" mains connection
6 "REMOTE" connection 13 "POWER" mains switch
7 Module for additional interfaces, optional:
Profibus, DeviceNet or Profinet
Product description