6.4 Reference value inputs
# Display Description Functions Data
type
Ac-
cess
type
Unit min. max. default
700 RUTimeSVal Set value run-up time 1 RW min 1 120 8
701 SpdSwPt1 Rotation speed switch
point 1
1 RW % 50 97 80
707 SpdSVal Set value in rotation speed
setting mode
2 RW % 20 100 65
710 Swoff BKP Backup pump switch-off
threshold for intermittent
operation
1 RW W 0 1000 0
711 SwOn BKP Backup pump switch-on
threshold for intermittent
operation
1 RW W 0 1000 0
717 StdbySVal Rotational speed set value
in stand-by operation
2 RW % 20 100 66.7
720 VentSpd Venting at rotation speed,
delayed venting
7 RW % 40 98 50
721 VentTime Venting time, delayed
venting
1 RW s 6 3600 3600
738 Gauge type Type of pressure gauge noGaug
TPR2xx
IKR2xx
PKR2xx
CMRx61
CMRx62
CMRx63
1 RW 0 6 0
777 PumpRotMax Confirmation of nominal
rotational speed
1 RW Hz 0 2000 777
794 Param. set Parameter set 0 = Basic pa-
rameter set
1 = Extended
parameter
set
7 RW 0 1 0
795 Servicelin Insert service line 7 RW 309
797 Address Device address 1 RW 1 255 1
Tbl. 17: Reference value inputs
Parameter set
34/56

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pfeiffer Vacuum TCP 350 and is the answer not in the manual?

Pfeiffer Vacuum TCP 350 Specifications

General IconGeneral
BrandPfeiffer Vacuum
ModelTCP 350
CategoryDC Drives
LanguageEnglish

Related product manuals