6.4 Reference value inputs
# Display Description Functions Data
type
Ac-
cess
type
Unit min. max. default
700 RUTimeSVal Set value run-up time 1 RW min 1 120 8
701 SpdSwPt1 Rotation speed switch
point 1
1 RW % 50 97 80
707 SpdSVal Set value in rotation speed
setting mode
2 RW % 20 100 65
710 Swoff BKP Backup pump switch-off
threshold for intermittent
operation
1 RW W 0 1000 0
711 SwOn BKP Backup pump switch-on
threshold for intermittent
operation
1 RW W 0 1000 0
717 StdbySVal Rotational speed set value
in stand-by operation
2 RW % 20 100 66.7
720 VentSpd Venting at rotation speed,
delayed venting
7 RW % 40 98 50
721 VentTime Venting time, delayed
venting
1 RW s 6 3600 3600
738 Gauge type Type of pressure gauge noGaug
TPR2xx
IKR2xx
PKR2xx
CMRx61
CMRx62
CMRx63
1 RW 0 6 0
777 PumpRotMax Confirmation of nominal
rotational speed
1 RW Hz 0 2000 777
794 Param. set Parameter set 0 = Basic pa-
rameter set
1 = Extended
parameter
set
7 RW 0 1 0
795 Servicelin Insert service line 7 RW 309
797 Address Device address 1 RW 1 255 1
Tbl. 17: Reference value inputs