# Display Description Functions Data
type
Ac-
cess
type
Unit min. max. de-
fault
023 Motor TMP Motor vacuum
pump
0 = off
1 = on
0 RW 0 1 0
025 OpMode BKP Backup pump
operating mode
0 = continuous opera-
tion
1 = intermittent opera-
tion
7 RW 0 1 0
026 OPMode TMP Rotation speed
setting mode
0 = off, final speed oper-
ation
1 = on, rotation speed
setting mode
7 RW 0 1 0
027 GasMode Gas mode 0 = heavy gases
1 = light gases
7 RW 0 1 0
028 Opmode Rem Operating mode
interfaces
0 = no priorities
1 = remote priority, with
pin 14 active high
7 RW 0 1 0
030 VentMode Venting mode 0 = delayed venting
1 = no venting
2 = direct venting
7 RW 0 2 0
035 Conf IO Accessory out-
put configura-
tion
0 = Heating
2 = Fan
7 RW 0 2 0
055 Conf AO1 Output AO1
configuration
0 = actual rotational
speed
1 = output
2 = current
7 RW 0 2 0
095
2)
RstCstVals Reset to factory
settings (relates
to actuating in-
structions and
setpoints)
no
yes
0 W - - no
Tbl. 15: Control commands
6.3 Status requests
# Display Description Func-
tions
Data
type
Ac-
cess
type
Unit min. max. de-
fault
300
3)
Remote priority 0 = no
1 = yes
0 R 0 1
301 Oil defic Operating fluid low
302
4)
Rotation speed switchpoint
reached
0 = no
1 = yes
0 R 0 1
303 Error code Error code 4 R
304
5)
Excess temperature, elec-
tronic drive unit
0 = no
1 = yes
0 R 0 1
2) Reset is only possible via the operator key on the unit
3) No text display of the parameter, instead symbol or key operated
4) No text display of the parameter, instead symbol or key operated
5) No text display of the parameter, instead symbol or key operated, detailed error description via parameter 303
Parameter set
32/56