Initializing the System for the First Time
Note: This is for the first time set-up only. In the on-screen directions you will be skipping steps
that would normally be required during consumable replacement.
Note: Some steps are applicable only to individual PHOENIX models. If your PHOENIX does not
have a UV lamp and/or Ultrafilter, these components will be greyed out on the Home Screen
Display.
• Ensure feed water supply is on and connected to system.
• Turn on power switch. Splash screen will display.
• Press START. Home Screen will display.
• After a 10 second delay, the pump will start and the system will begin the initial rinse cycle.
TO ACTIVATE THE ULTRAFILTER FOR BIOCHEM AND GENOME Models:
• Press and hold the ULTRA FILTER icon and press NEXT to initialize the Ultra Filter.
• Follow the instructions displayed.
o Skip on-screen instruction for Steps 5 and 6 as the ultrafilter is already installed.
o During Step 9, the purging process may take several minutes.
o Complete the procedure by pressing RESTART to return to the Home Screen.
TO ACTIVATE THE UV LAMP FOR CITATION and GENOME Models:
• From the Home Screen, press and hold the UV LAMP icon. Press NEXT to continue.
o Follow instructions displayed. It will not be necessary to replace lamp.
o Complete the procedure by pressing RESTART to return to the Home Screen.
TO ACTIVATE THE FILTER PACKS FOR ALL MODELS:
• From the Home Screen, press and hold the CONDITIONING icon. Press NEXT to continue.
o Follow instructions displayed. As this is the first time initialization process a new
pack should have been installed. It will not be necessary to replace the pack as
specified in the instructions.
o Complete the procedure by pressing RESTART to return to the Home Screen.
• From the Home Screen, press and hold the POLISHING icon. Press NEXT to continue.
o Follow instructions displayed. As this is the first time initialization process a new
filter should have been installed. It will not be necessary to replace the filter as
specified in the instructions.