28
CITATION and GENOME:
UV lamp life is estimated at 2 years of normal use from date of installation. The UV lamp health
meter on the HOME SCREEN displays the proportional remaining life in this 2 year period as shown
below:
A replacement alarm will be displayed at 1 month remaining. It is recommended that the UV lamp
be replaced at that time to maintain low TOC levels.
GENOME Model:
Ultrafilter life is estimated at 1 year of normal use from date of installation. The Ultrafilter health
meter on the HOME SCREEN displays the proportional remaining life over this time period as
shown below:
A replacement alarm will be displayed at 2 months remaining. It is recommended the Ultrafilter be
replaced at that time to maintain acceptable bioburden.
Adjusting Resistivity Alarm Set Points
The point at which the conditioning pack and polishing filter transition from green health status to yellow
can be adjusted. Note: Software allows for a 2 MΩ range above the set point in which to change the health
meter display. Always make the set point 2MΩ below the desired display transition resistance. Maximum
set point allowed for conditioning is 10 MΩ and 14 MΩ for polishing.
To adjust the set points:
• Home Screen – Press Menu
• Main Menu – Press Configuration
• Configuration Menu – Press Alarm Setup
• Press appropriate field box; key pad will open
• Adjust values as required
• Press ENT to exit key pad
• Press and hold SAVE until beep is heard
• Press Home icon to return to Home Screen