Installation
Installation Requirements
Power Requirements:
o -1” Option - 120 VAC/ 60 Hz @ 1.0 amp
o -2” Option – 220VAC / 50 Hz @ 0.5 amp
GFCI protected outlet within:
o 6 feet for system
WARNING: Must be a grounded electrical connection.
Feed Water Supply: Water source meeting minimum requirements as shown
below and located within 3 feet of the system.
Parameter
Requirement
Source/Type
Reverse osmosis (RO) or Service deionization (SDI) preferred. 0.2µm
particulate prefiltering is recommended for SDI and tap water feeds.
Conductivity
RO: < 20µS/cm
SDI: > 1 MΩ (Resistivity)
TOC (Using appropriate
conditioning as necessary)
RO: < 50 ppb
SDI: < 200 ppb
Temperature
20° F to 100°F (5° C to 38°C)
Pressure
20 psig minimum/90 psig maximum (1.38 to 6.21 bar)
Fouling Index
Silt Density Index: < 3
Dissolved Carbon Dioxide
< 30 ppm
Free Chlorine
< .05 ppm
An upstream isolation shut off valve is highly recommended to facilitate future
servicing
The Phoenix is supplied with an inlet connection for 3/8” OD tubing
The Phoenix is supplied with an outlet drain port for 3/8” OD tubing
What you will need:
#2 Phillips Head Screw Driver
Teflon Tape
Upstream Isolation Valve (Recommended)
Basic Hand Tools (Adjustable wrench)
What you’ll find in the box:
Phoenix System
Certificates of Conformance
Instructions
Power cord
Final Filter (Point of Use Filter)
Drain Fitting
6’ Length of 3/8” Diameter Inlet Tubing & Drain Tubing

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Phoenix ACADEMY and is the answer not in the manual?

Phoenix ACADEMY Specifications

General IconGeneral
BrandPhoenix
ModelACADEMY
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals