Installation Requirements
• Power Requirements:
o “-1” Option - 120 VAC/ 60 Hz @ 1.0 amp
o “-2” Option – 220VAC / 50 Hz @ 0.5 amp
• GFCI protected outlet within:
o 6 feet for system
WARNING: Must be a grounded electrical connection.
• Feed Water Supply: Water source meeting minimum requirements as shown
below and located within 3 feet of the system.
Reverse osmosis (RO) or Service deionization (SDI) preferred. 0.2µm
particulate prefiltering is recommended for SDI and tap water feeds.
RO: < 20µS/cm
SDI: > 1 MΩ (Resistivity)
TOC (Using appropriate
conditioning as necessary)
RO: < 50 ppb
SDI: < 200 ppb
20° F to 100°F (5° C to 38°C)
20 psig minimum/90 psig maximum (1.38 to 6.21 bar)
• An upstream isolation shut off valve is highly recommended to facilitate future
servicing
• The Phoenix is supplied with an inlet connection for 3/8” OD tubing
• The Phoenix is supplied with an outlet drain port for 3/8” OD tubing
What you will need:
• #2 Phillips Head Screw Driver
• Teflon Tape
• Upstream Isolation Valve (Recommended)
• Basic Hand Tools (Adjustable wrench)
What you’ll find in the box:
• Phoenix System
• Certificates of Conformance
• Instructions
• Power cord
• Final Filter (Point of Use Filter)
• Drain Fitting
• 6’ Length of 3/8” Diameter Inlet Tubing & Drain Tubing