SYMPTOM
CAUSE
REMEDY
Unit does not rinse-up to
desired resistivity.
New units or cartridge
replacements require more
time to rinse-up.
Allow more time when installing
or replacing media.
Use a hand-held resistivity meter
to check that the resistivity is
increasing over time
Air is getting trapped in the
system.
The system is open to
atmosphere.
Bleed air at high points in
plumbing system and/or filter
housing. Use vent valve supplied
where applicable.
Excessive flow, turbulence,
and/or cavitation can cause
gases to come out of solution.
Check flow and velocity
throughout system to ensure it
does not exceed filter
requirements or industry
standards.
The point of use 0.2 micron
filter is bound with air.
Bleed the filter by opening the
bleed valve and allowing water
to flow until all air is expelled.
The 0.05 micron filter is
bound with air. (BIOCHEM
and GENOME)
Bleed the filter by opening the
bleed valve and allowing
water to flow until all air is
expelled.