PORTER Reliant
10
2.8 STANDARD ACCESSORIES
To make the warranty active it is necessary that a copy of the supplied Unit Passport is sent, through the
agent, to the manufacturer; for want of this the warranty will decline.
2.9 TECHNICAL SPECIFICATIONS
Chamber dimensions
Ø = 240 mm
Depth = 384 mm
Auto-switching-off
elapsed 30’ from the end of a
cycle and without any action
Chamber capacity
17,5 l
Dual water tank
3,5 liters each (used and
clean water tanks)
Maximum load
4 kg (solid)
1,5 kg (porous)
Bacterial filter
0.3 µm al 99.97 %
Warming-up time
20’ from room temperature
10’ with pre-heated chamber
Clock battery
(may be replaced only by au-
thorized service)
Varta CR2032
Sterilization time
from 3’ to 90’ depending on the
selected cycle
Transmitted heat
in environment
0,18 J/h at 23°C
Drying time
from 6’ to 30’ depending on the
selected cycle
Sound emission
61 dB/A at 1 m
External dimensions
443 x 570 x 428 mm
(L x D x H)
Working cycle
continuous
Net weight
49 Kg
Pollution grade
2
Power supply voltage
120V +/-10%
Transient overvoltage
II
Frequency
60Hz
Max consumption
1500 W
Maximum volume available
on the trays
10 l
Average consumption
600 W
Maximum chamber
temperature
135°C (-0/+2°C)
Standby consumption
1 W
Safety valve
working pressure
2,55 bar
AC Fuses
2 x 15AT 250V (type 6.3x32 )
UL listed
2.9.1 Environment operating conditions
The sterilizer is designed to operate in environments at temperatures between 3 °C and 40 °C, relative
humidity not greater than 95%, atmospheric pressure from 750 mBar to 1050 mBar and altitude between
0 and +2500 meters.
NEVER USE IN PRESENCE OF INFLAMMABILE ANAESTHETIC GAS