PORTER Reliant
12
3.2 GETTING STARTED
These operations should be carried out only by qualified personnel, wrong settings could
affect the quality of the sterilization.
Check the electrical requirements and connect the power supply cable to the mains socket.
1. The equipment is delivered with empty tank and is therefore
necessary to fill it completely with demineralized water. Fill the
tank without exceeding the maximum level reference marked on
the tank opening.
The use of low quality demineralized water could create scale
deposits on the instruments, inside the chamber and on the
trays. Carefully read the label of the demineralized water con-
tainer. Do not use tap water even if treated with filters or sweet-
eners.
Do not use water for batteries, other liquids or additives
which can cause irreversible damage to the equipment and
be a risk for the operator.
2. Turn on the equipment with the main switch on the back panel.
This switch can be left ON since the electrical consumption in
stand-by is almost nil.
3. Remove the basket and the trays from the chamber and close
the door.
When the unit is off, the door is blocked; if the block persists, turn
off and then on again the unit.
4. Hold on the key
and push the key
Power
; the display
shows the message <SET ALT 100 MT> with the set altitude
value (100 m).
Use both keys to adjust the value as necessary according to the
actual site altitude (see next page).
Then push the key
Set
to confirm the displayed value and start
the automatic initialization procedure with the entry of water in
the hydraulic circuit and chamber.