6. MAINTENANCE
6.1 AUTOMATIC PERIODIC CLEANING CYCLE
For a proper operation of your sterilizer it is indispensable that a correct and regular maintenance is car-
ried out. To this end it is important to perform a cleaning procedure, as shown on the following, at least
once every 60 cycles.
For further safety, after 60 cycles carried out without an intermediate cleaning cycle, the equipment
will display the message <NEED CLEANING >.
IMPORTANT
Take the basket and the trays out of the chamber and wash with an ordinary dish washing powder,
rinse with water and wipe.
DO NOT USE ABRASIVE PRODUCTS.
WARNING
DO NOT PERFORM THE CLEANING CYCLE WITH TRAYS IN THE CHAMBER.
Carry out the chamber surface cleaning when the equipment is cold.
NOTE
To open the door and carry out these preliminary operations, it is necessary to switch on the
equipment; ended the maintenance, switch off the unit to avoid excessive heating. To close the
door switch on the unit again.
1. Drain the main tank by means the supplied drain tube fit into the white
fast fitting.
2. Melt one cleaning tab in 1-liter solution of demineralized water and
pour it into the main tank.
3. Close the door and put the unit in Stand-by by means of the key
POWER (message OFF on the display).
4. hold pushed the key START/STOP and push the key POWER to start
the automatic cleaning cycle. The display is showing CLEANING.
5. The cleaning cycle takes between 10 and 20 minutes, and cannot be
interrupted (command panel keys disabled).
6. Over the cycle, the indication Led MAX and MIN will blink.
7. Open the door and drain both tanks again. During this phase the dis-
play is showing the message CLEANING. Over the drain, remove the
tube from the fast fitting.
8. Push the key START/STOP: the water pump will start for a few sec-
onds to empty completely the hydraulic circuit.
9. During this time the door must be left open. Carry out the main tank fill
with demineralized water; on turning off the indication Led MIN, the
water pump starts for a few seconds in order to fill the internal hydrau-
lic tubing.
10. The end of the cleaning cycle is signalled with the message READY
on the display.
11. In case the equipment will work using the recycled water, wipe the
chamber before starting a new sterilization cycle; on the contrary it is
sufficient to close the door and the water in the chamber will be col-
lected into the recovery tank.
12. Close the door and switch off by the key POWER.
CAUTION! The water contained in the recovery tank is to be regarded as biologically contaminating,
therefore when this tank is emptied, suitable precautions should be taken. The disposal of recovery water
needs to be done in accordance with the national orlocal regulation.
IMPORTANT
The sterilizer operates with a closed hydraulic circuit, therefore electro-valves, pump and tubing
could be progressively clogged by particles and residues from the sterilization process. So it is
very important to carry out regularly the cleaning cycle as above indicated.