PORTER Reliant
20
5.3 SPECIAL CYCLE
The operator can set a customized sterilization cycle as follows:
Push the key
Set
and
S
in a sequence
The display Pressure shows:
SET TEMP
Set the process temperature value between 105
and 135°C by the keys and
The display Pressure shows:
SET TIME
Set the process time between 3 e 90 minutes by
the keys and
The display Pressure shows:
SET DRY
By the keys and
set the dry-
ing time between the pre-set available value of
5, 10, 20 and 30 minutes; the value is shown
close to message DRY.
Push again the key
Set
to exit the programming mode.
The parameter values for the SPECIAL cycle are automatically stored and maintained until new values are
set through the same procedure.
The combination of time/temperature parameters set by the operator can lead to cycles that do
no assure the sterilization. It is necessary to check the effectiveness of the customized cycle by
means of suitable tests.
At the end of the SPECIAL cycle only the Led READY will switch on - not STERILIZE - to signal
that the efficiency of the cycle selected by the operator has not been validated by the manufac-
turer.