PORTER Reliant
17
Opening the door, the displays will show again the current time, chamber temperature and pressure, and
the sterilizer is ready for a new cycle.
If a printer is connected and ready, a report will be issued during the cycle phases with the more signifi-
cant data; the report can be filed as proof of the sterilization process performed.
The operator can arrange other load on the trays and start a new sterilization cycle, with the advantage of
shorter heating-up time as the chamber is already warm, or press key Power to turn off the unit (stand-
by).
If the door is not opened or a key pressed within 30 minutes, the unit switches to stand-by (OFF)
automatically.
Should any failure or error occur during the cycle, Led ALARM turns ON and the display Time will show
alarm type and code (see chapter ALARM).
4.3 STOPPING THE CYCLE
To stop the sterilization cycle, press key
Start/Stop
. The display Time shows the message “MANU
STOP”.
Before opening the door, make sure that the display Pressure is showing the value 0. A safety device will
anyway prevent from opening the door with the chamber over-pressurized. To unlock the door press key
Start/Stop
.
Remove the load and check for the presence of water into the chamber. In case of wrapped instruments,
we suggest to replace with new bags.
Before loading the chamber again, dry it carefully and wait 10 minutes to allow the water to evaporate and
be drained completely.