PORTER Reliant
11
3. INSTALLATION
3.1 BASIC REQUIREMENTS
1. Check that the mains voltage of your electrical installation matches the
value indicated on the equipment plate, the electrical socket is capable
to supply at least 15A and is provided with an efficient earth connec-
tion. In case the installation makes inaccessible the power supply
switch, provide for a proper electrical breaker.
The manufacturer will not be responsible for damages to people
or things caused by an unsuitable electrical installation or miss-
ing of the ground connection.
2. The equipment must be installed on a flat surface; adjust
the front feet to have a slight slope and so make the out-
flow of water easier during the drain phase.
WARNING: do not place the unit on surfaces which could
cause fire or fume if hot items fall from the equipment
3. For the correct operation, it is mandatory to leave a free
space of at least 4 cm on the rear side of the equipment.
4. Do not install the equipment near heat sources, in humid
or not well aired environment; in a sterilization room is re-
quired a minimum of 10 air changes per hour, a recirculat-
ing ventilation system can not be used.
5. On the rear panel it is located the safety valve, if it oper-
ates for overpressure it may be released hot steam: locate
the unit to eliminate the risk of burning for the operator
(f.e. near a wall)
The sterilizer can also be installed in a cabinet. In this case,
provide adequate space (>10 cm) for aeration.