Although the sterilizer’s average life is about 8-12 years, it is compulsory to carry out periodic controls in order
to check the correct calibration and for possible wear of components. The control interval, as reported on the
Service Book, is 1 Year or 1500 cycles for the ordinary maintenance, and 4 Years or 10.000 cycles for the
special maintenance. The sterilizer is configured to signal on the display, as the first term comes to expire,
the message “ NEED SERVICE”. This feature is in compliance with the norm EN13060 and answers to spe-
cific safety requirements. The maintenance activities must be performed by authorized technicians ( provided
with card or certificate released by Porter) and reported on this Service Book.
Model .......................................................... Serial Number ........................
Installation Date ............................................ Altitude ..................................
Installer ...............................................................................................................
Reseller ...............................................................................................................
Sterilization Manager: .......................................................................................... .
Standard periodic checks
*) To be carried out in Porter service department (service kit code 118005 is composed by: safety
valve, closing device, bacterial filter, gasket door, 3 net filter, OR for water filter)
In case of direct ship or consign for shipping the sterilizer for service center or factory repairing, include a copy of
the filled-in Service Book pages.