- iv -
CAUTION
This instrument is shipped with a grounding type power cable.
To reduce the risk of electric shock, always plug the cable into a
grounded power outlet.
CAUTION
Wipe the forehead rest with ethanol or glutaraldehyde solution to
disinfect it each time a different examinee uses it, in order to
prevent infection.
CAUTION
Do not use the device near, on, or under other electronic equipment
to avoid electromagnetic interference with the operation of the
device.
CAUTION
Vulnerable to electrostatic non-conformity points and may cause
screen shake
CAUTION
Do not use the device in the same room with other equipment such
as life-support equipment, other equipment that has major affects
on the life of patient an results of treatment, or other measurement
or treatment equipment that involve small electric current.
CAUTION
Do not use the device simultaneously with portable and mobile
radio frequency communication systems because it may have an
adverse effect on operation of the device.
CAUTION
Do not use cables and accessories that are not specified for the
device because that may increase the emission of electromagnetic
waves from the device or the system and decrease the immunity of
the device to electromagnetic disturbance.
CAUTION
Do not position the equipment to make it difficult to operate the
disconnection device. (Appliance coupler or separable plug)
CAUTION
Do not placed the multiple socket-outlet for PLC-8000(pola) system
on the floor in order to prevent liquid penetration and damage to the
product.
CAUTION
PLC-8000(pola) system shall not be connected with additional
multiple socket-outlets or extension cords in addition to a
designated single multiple socket-outlet.
CAUTION
Maximum permissible load of each socket-outlet used for the PLC-
8000(pola) system, shall not be less than 100VA.
CAUTION
If non-medical electrical equipments (e.g. Video monitor, IT
equipment, etc.) that are connected with the PLC-8000(pola) are,
directly connected to the wall socket-outlets, high touch current may
flow since the earth continuity is not ensured.
CAUTION
Multiple socket-outlet should be a grounding-type and complied with
IEC 60884-1.