12 PRODIGIT
2-1. Installation and removal of 3310F series plug-in load module
The 3310F series Electronic load module operates from within the 3300F/3302F/3305F mainframe.
The mainframe is required to provide power to the module’s control circuitry. It is also needed for the
computer interfaces, analogue programming input and the 150 store/recall memory.
Unless the 3300F/3302F/3305F mainframe and 3310F series Electronic load module were purchased
separately, the 3310F series Electronic load module should be installed in the 3300F mainframe
before shipment from Prodigit.
One of the benefits of the modular approach is that different models of load module can be operated
from within the same mainframe. It is easy for the user to reconfigure the mainframes by changing or
adding different load modules.
The following procedure should be followed for installing or removing the 3310F series load module in
or out from the 3300F /3302F/3305F.
2.1.1. Installation of 3310F series plug-in load:
2.1.1.1 Turn the 3300F/3302F/3305F mainframe power OFF before inserting the
3310F series load module. Failure to switch the mains power off may result in
Damage to the plug-in module’s circuitry.
2.1.1.2 Align the upper and lower grooves of the 3300F mainframe with the upper and
Lower guides of the selected compartment.
2.1.1.3 If correctly positioned the 3310F series load module will slide in easily until
Some 30-40mm is left protruding from the mainframe. At this point a little more
Force will be required to seat the load module’s circuit board in the
Interconnecting jack of the mainframe. It is recommended that the binding posts
On the load module’s front panel be used to push the module home.
2.1.1.4 Use the supplied screw to fasten the load module to the mainframe. The
Screw hole is located at the end of the pull out handle at the bottom right hand
Corner of the 3310F load module. The screw location is shown on Fig 2-1 and
Is below and to the right of the LOAD + binding post.
2.1.1.5 Only after all the load modules are installed to the 3300F/3302F/3305F
Mainframe should the mains power be switched ON.
2.1.2. Removal of 3310F series plug-in load:
2.1.2.1. Firstly ensure that the mains power to the 3300F/3302F/3305F mainframe is
Switched off. Failure to do so may result in damage to the load module.
2.1.2.2. Take the screw out of the pull out handle in the lower right corner of the module.
2.1.2.3. After removal of the screw the handle can be pulled towards you to lever the
Module out of the mainframe.