3310F Series Operation Manual 75
4-13 SHORT testing
4.13.1. SHORT Manual control
Example:
4.13.1.1. Setting SHORT test, press Short key to the next step.
4.13.1.2. Press UP key, setting Short time to 10000ms, press Short key to
the next Step.
4.13.1.3. Press down key, setting V-Hi voltage to 1.000V, press Short key
to the next Step.
4.13.1.4. Press down key, setting V-Lo voltage to 0V, press Short key to the
next Step.
4.13.1.5. Press START/STOP test key.
4.13.1.6. Short test finish, the UUT’s drop voltage is between V_Hi and
V_Lo limitation, then middle 5 digits LCD display will shows
"PASS"