ProMinent DULCOMETER D1CA - pH;ORP SN6 Socket Connections

ProMinent DULCOMETER D1CA
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
pH/Redox über SN6-Eingang / pH/ORP via SN6 socket
Pt 100 (temperature)***
Mains power Mains power
Netz
Extern Pumpe 2 senken
Extern Pumpe 1 heben
Normsignal-Ausgang 1
0/4-20mA
Standard signal output 1
0/4-20mA
Netz
external
Extern
(potentialfrei)
(potentialfrei)
with
without
SN6-input
Pt 100 (Temperatur)***
87
65 43
21
pH/Redox (ORP)
SN6-Eingang
ohne/
mit/
Potentialausgleich
Liquid reference potential
Klemmenanschluss für pH / Redox über SN6-Eingang
Terminal connection for pH / ORP via SN6 socket
X2
1211109
1211109
+
-
pH/Redox (ORP)
**
Temperature
input (Pt 100)
Temperatur-
Eingang (Pt 100)
Frequency outputs
(pumps)
Frequenz-Ausgänge
(Pumpen)
i. e. recorder
z. B. Schreiber
internal
Intern
Liquid reference potential
SN6
Potentialausgleich
SN6
pH/Redox (ORP)
*
*
clamp terminals X2.12 and X2.11 are connected internaly.
* Bei Version D (Schalttafeleinbau) sind
die Klemmen X2.12 und X2.11 intern verdrahtet.
* Version D (Control panel installation) only:
External pump 2 decrease (potencial free)
External pump 1 increase (potencial free)
*** Correction variable for pH (see also p. 53)
*** Korrekturgröße für pH (vgl. auch S. 53)
** ACHTUNG: Bei Anschluss der Sonde ohne Potenzialausgleich
die Klemmen X2.9 und X2.10 brücken.
** IMPORTANT: When connecting the electrode without liquid reference potential
clamp terminal X2.9 and X2.10 have to be closed.