20
Changing, Storing, and Recalling Custom Setups
Measurement Mode
The measurement mode is selected by pressing one of the four keys in the
PARAMETER group of keys on the front panel.
AUTO In this mode, the meter automatically selects the most appropriate
measurement on the device. The selection is made according to the following
criteria:
If |Q| < 0.125, the meter selects the R+Q measurement mode.
If Q > +0.125, the meter selects the L+Q measurement mode.
If Q < 0.125 and the meter is in the series equivalent circuit mode, the
C+R measurement mode is selected.
If Q < 0.125 and the meter is in the parallel equivalent circuit mode, the
C+D measurement mode is selected.
R+Q In this mode, resistance is displayed on the main parameter display and the
Q (quality factor) is displayed on the secondary parameter display. The R-value
is either the series or parallel equivalent resistance of the component under test.
The unit for R is , k, or M. The Q value is the dimensionless ratio of the
imaginary part of the component’s impedance to its real part. If Q has a positive
(+) value, the reactive component of the device under test is inductive. If Q has a
negative () value, it is capacitive.
L+Q In this mode, the series or parallel equivalent inductance value of a
component under test is displayed on the main parameter display units in units of
µH, mH, or H, and its dimensionless Q value is displayed on the secondary
parameter display.
C+D In this mode, the series or parallel equivalent capacitance value of a
component under test is displayed on the main parameter display in units of pF,
nF, or µF. On the secondary parameter display, the dimensionless ratio D 1/Q
is given. D is normally a small value for a good capacitor.
C+R In this mode, the capacitance is displayed on the main parameter
display and the equivalent resistance is displayed on the secondary parameter
display. The unit for resistance is unless the k indicator LED is lit.
Frequency
The output frequency is accurate to 100ppm (0.01%). Frequencies are set by
pressing the up/down keys () on the front panel keypad until the LED
corresponding to the desired frequency is illuminated. If the meter is on range
hold on Range 0, the 100 kHz frequency selection is disabled. If 100 kHz is
selected while in Range 0, the meter will beep and “r-f error” (range frequency
error) will be appear on the alphanumeric display. For more details, see the
section on Range.