EasyManuals Logo

Protek 9216A User Manual

Protek 9216A
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
31
such as taking a measurement, or auto calibration, take a long time to execute and there
is no other way of determining when they are done. The NO COMMAND bit may be read
while commands are being executed by doing a serial poll. There is no way to read this
bit over RS232. Note that using the *STB? query to read this bit will always return the
value 0 because it will always return an answer while a command is executing - the
*STB? command itself.
Definitions of Status Bytes
Serial Polling
bit
name
usage
0
Ready
The 9216A is ready to perform a measurement
1
Not used
2
Not used
3
LCR
An unmasked bit in the LCR status register has been set.
4
ESB
An unmasked bit in the standard status byte has been set.
5
RQS / MSS
SRQ (Service Request) bit.
6
No command
There are no unexecuted commands in the input queue.
Standard Event
bit
name
usage
0
OPC
Set by the OPC command when all measurements are complete
1
Not used
2
Query Error
Set on output queue overflow (too many responses waiting to be transmitted.)
3
Not used
4
Execution err
Set by an out of range parameter, or non-completion of some command due a
condition such as an incorrect operating mode.
5
Command err
Set by a command syntax error, or unrecognized command
6
URQ
Set by any key press
7
PON
Set by power on

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Protek 9216A and is the answer not in the manual?

Protek 9216A Specifications

General IconGeneral
BrandProtek
Model9216A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals