EasyManuals Logo

Protek 9216A User Manual

Protek 9216A
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
16
1. Set the 9216A to 1 kHz, 1 V and constant voltage. Set the scope to 1 V/div verti-
cal and 0.5 ms/div horizontal. Connect a 10 probe to the scope.
2. Place the tip of the probe into the "+" side of the fixture and connect the ground
clip to the center guard.
3. The scope should display a sine wave that occupies two divisions horizontally
and about 5.5 divisions peak to peak vertically (1.0 Vrms - 2.83 V peak to peak).
There should be no irregularities in the waveform.
4. Change the amplitude setting of the 9216A to 0.25 and 0.10 V in succession and
verify that the output is within 2% of nominal.
5. Set the amplitude back to 1.0 V. Change the 9216A and scope settings to verify
that the output at 100 Hz, 120 Hz, 10 kHz and 100 kHz is within 2% of nominal.
Resistance Measurement
This test verifies that the 9216A operates and is able to measure a component in each of
its ranges. The readings obtained should be within (tolerance of the component + toler-
ance of the 9216A).
1. Press the key sequence RCL 0 ENTER to put the unit in its default setup.
2. Perform open and short circuit calibrations for the fixture configuration to be used.
See Volume 1 for details on these null calibrations.
3. Set the unit to the R+Q measurement mode, series equivalent circuit, and 1 kHz
test frequency. Install the 24.9 resistor.
4. Verify that the meter reads the resistance correctly to within 0.15%. Verify that
Q is a small value, about +0.0001 or smaller. Install the 402 resistor. Verify
that the meter reads the resistance correctly to within 0.15%. Verify that Q is a
small value, about +0.0001 or smaller.
5. Change the equivalent circuit to parallel. Install the 6.34 k resistor. Verify that
the meter reads the resistance correctly to within 0.15%. Verify that Q is a small
value, about 0.0001 or smaller.
6. Install the 100 k resistor. Verify that the meter reads the resistance correctly to
within 0.15%. Verify that Q is a small value, about 0.0002 or smaller.
Capacitance Measurement
This test verifies that the 9216A is able to measure components at different frequen-
cies. The limits of the readings are the same as before: (component tolerance + meter
tolerance).
1. If the fixture configuration has changed, perform open and short circuit calibration.
2. Set the 9216A to the C+D measurement mode, parallel equivalent circuit, and 1
kHz test frequency.
3. Install the 22 nF capacitor. Verify that the unit reads the capacitance correctly to
within 1.10%. Verify that D is below 0.0001.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Protek 9216A and is the answer not in the manual?

Protek 9216A Specifications

General IconGeneral
BrandProtek
Model9216A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals