38
Summary of Binning Setups
Pass/Fail Setup
Enter the nominal value and limits for Bin 0. Enter the QDR fail value for Bin 8. Make
sure no other bins are open (set their limits to zero). Parts that pass fall into Bin 0; and all
other parts fall into Bin 8 or Bin 9.
Nested or Overlapping Bins
Enter the nominal value and limits for Bin 0. For subsequent bins, enter only the limits,
making sure the tighter tolerance parts use the lower bins. If the limits are symmetrical,
only enter the upper limit. Enter the QDR limit for Bin 8. Make sure all other bins are
closed. Parts that pass fall into one of the pass bins. Parts that fail the QDR test fall in
Bin 8 and parts that fall into no other bin fall into Bin 9.
Sequential Bins
For sequential bins with a single nominal value, follow the same procedure as for nested
bins mentioned above.
For sequential bins with multiple nominal values, enter the nominal value and limits
for each open bin. If the limits are symmetrical, only enter the upper limit. Enter the QDR
fail value for Bin 8. Make sure that unused bins are closed (limits set to zero) and there
are no unwanted gaps between bins.
Parts that pass fall into one of the pass bins. Parts that fail the QDR test fall in Bin 8
and parts that fall into no other bin fall into Bin 9.