8
listed on the packing slip are included. If any items are missing or damaged, please
contact Pyxis Customer Service at service@pyxis-lab.com.
1.3 Standard Accessories
• Sample Vials - two 10 ml, round, 0.78 inch (20 mm) pathlength, glass vials,
which can be used for all measurements including turbidity and fluorescence
measurements
• 4 AA alkaline batteries
• Instrument Manual, also available from www.pyxis-lab.com
1.4 Optional Accessories
• 25 ml sample vial
• 100 ppb PTSA standard in a 500 ml brown plastic bottle
• 50, 250,and 500 ppb fluorescein standards in a 500 ml brown plastic bottle
1.5 Sample Vial Compartment
The sample vial compartment is shown in Figure 1, along with a 10-ml sample vial.
When the sample vial is inserted into the sample vial compartment, the triangular mark
on the sample vial should be aligned approximately with the 6 o’clock position of the
sample vial compartment or any position consistently.
The sample vial compartment should be kept clean. A small amount foreign material
could significantly affect turbidity and fluorescence measurement results. Use a soft
cloth or lint free paper tissue to clean sample vial compartment periodically. Remove
debris, scale, and deposit promptly.
1.6 Light Shield Cover
The light shield cover is shown in Figure 2. The light shield cover can be conveniently
slid between the open and closed positions. The light shield cover is held firmly at the
rest positions by permanent magnets.
The light shield cover should be in the closed position during storage, transportation,
and measurements, especially during the turbidity and fluorescence measurements.
When turned on, The SP-910 carries out self-diagnosis including checking the
performance of a variety of optical devices. The light shield door shall be at the closed
position to shield interference from ambient light during self-diagnosis.
Care should be taken to avoid water or debris being trapped in the track of the light
shield door.