6
General Description
Specification
1
Colorimeter Wavelength: 365/420/455/525/560/570/630 nm
Turbidity Excitation: White/Infrared LED/90 degree scattering
Fluorescence Excitation Wavelength: 365/460 nm LED
Fluorescence Emission Wavelength: 410/520 nm
Wavelength Accuracy: ±1 nm
Absorbance Reproducibility: 0.005 au (0 - 1.0 au) (3 sigma)
Absorbance Linearity Range: 0 - 1.0 au
PTSA Reproducibility: 1 ppb PTSA (3 sigma)
PTSA Detection Limit: 1 ppb
PTSA Range: 0 - 300 ppb
Fluorescein Reproducibility: 0.2 ppb or 2% of the value
Fluorescein Detection Limit: 0.1 ppb
Fluorescein Range: 600 ppb
Turbidity Reproducibility: 1 NTU (3 sigma)
Turbidity Detection Limit: 1 NTU
Turbidity Range: 1 - 200 NTU
Battery: 4 AA alkaline
Typical Battery Life: 3 months
Display: Graphical LCD 192x128 pixels,
visible under direct sunlight
Instrument Dimension: L 265mm W 88mm H 62mm
Instrument Weight: 600 g without batteries
Storage Temperature Range: 0 - 140°F (-18 - 60°C )
Operation Temperature Range: 40 - 120 °F (4 - 49°C )
Humidity: 85% at 106 °F (41 °C )
Environmental: IP67, dustproof and waterproof
Note:
1. Specifications are subject to change without notice for continuous development.
2. The fluorescein range in earlier versions of the SP-910 may be only up to 20 ppb.
To extend the upper limit to 600 ppb, please contact Pyxis customer support at
service@pyxis-lab.com.