EasyManuals Logo

RADIOMETER FLEX ABL90 User Manual

RADIOMETER FLEX ABL90
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
11
Capillary tube samples – preanalytical errors, continued
Sample collection and handling
5. Sample should be free of air bubbles
• Air bubbles can affect O
2
and CO
2
values
• Ensure the end caps match the type of capillary tube
• Cap both ends of the sample to avoid contact with room air
6. Mixing
• Mix GENTLY with the mixing wire and magnet to avoid hemolysis
of the sample
• Hemolysis may cause K
+
, Na
+
and Ca
2+
• Mix 20 times along the full length of the capillary tube or mix
manually by inverting the capillary tube 20 times
7. Analyze quickly
• If not analyzed quickly, the sample will continue to metabolize, causing
the following:
pO
2
since O
2
will still be consumed
pCO
2
since CO
2
will still be produced
pH due to change in CO
2
cCa
2+
the change in pH will influence the binding of Ca
2+
to protein
cGlu since glucose will be metabolized
cLac due to metabolism
If not analyzed <30 minutes, store plastic capillary tubes at room temperature. Glass capillary tubes
can be stored in ice slurry.
11 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals