Replacements
22
Consumables Default tests or Recommended replacement
activities per day interval after installation
Solution pack 10 Maximum 30 days or when the
number of activities is zero
Sensor cassette 10 Maximum 30 days or when the
number of tests is zero
Inlet gasket holder 10 12 months
Inlet connector gasket 10 12 months
Storage of consumables
Sensor cassette – storage temperature 2-8 °C/35-46 °F
Solution pack – storage temperature 2-25 °C/35-77 °F
Always ensure stock rotation.
Replacements
Replacement intervals for consumables and inlet connector gasket
The recommended replacement intervals shown in the table below are only meant as a
guideline. They are based on a default of 10 sample analyses per day. For analyzers with a
higher sample throughput, the number of expected measurements per day can be changed
in the setup, so the analyzer can calculate the most probable replacement date and send a
message about it.