EasyManuals Logo

RADIOMETER FLEX ABL90 User Manual

RADIOMETER FLEX ABL90
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Changing consumables continued
Changing consumables – sensor cassette
Note: The analyzer has to condition a new sensor cassette after it is
installed. This process can take up to 4 hours.
• Show on the analyzer screen how to enter Analyzer Status
and find > Consumables > Replace > Replace sensor cassette
• Show how to start the video guidance and wait until the sensor
cassette compartment opens
• Let the trainee try to remove the sensor cassette and dispose of it
as biohazardous waste according to local guidelines
• Tell that actions have to be confirmed at the analyzer to advance
the video guidance, so press > Action Completed
Let the trainee do all of the following steps:
• Pull the foil off the new sensor cassette pack
• Unscrew the lid and lift out the sensor cassette
• Observe that the trainee tap > Action Completed before putting
the new sensor cassette in place
• Follow the on-screen guidance for completing the procedure
Note: If you tap the “Exit conditioning” button, the analyzer will not
perform the more frequent calibrations that normally follow in the 24
hours after a sensor cassette replacement. This may have a negative
effect on the accuracy and precision of patient results during this
time.
18
18 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals