Overview of quality control management
Quality control management is important as it evaluates the performance of the
analyzer to make sure that the patient results are accurate and precise.
The analyzer manages quality control automatically, but if local, federal or state regulations
require additional quality control procedures, operators can do them.
Automatic quality control management (AQM) is the name given to quality control
procedures that the analyzer is programmed to do automatically.
About built-in QC measurements
The analyzer uses the three levels of QC solution contained in the solution pack to do
built-in QC measurements. These QC solutions are automatically registered in slots A,
B and C when a solution pack is installed.
Note: The solution in slot A is S9030, the solution in slot B is S9040 and the solution in
slot C is S9050.
Built-in QC measurement frequency
A built-in QC measurement is scheduled by default to be done every 8 hours. One
measurement a day is done with each QC solution. Built-in QC measurements are also
scheduled by default to be done in connection with these activities:
• Replacement of the solution pack
• Replacement of the sensor cassette
Startup
You can edit the schedule for built-in QC measurements.
Quality control and calibrations
26
26 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals