Calibration results
• A calibration result is a combination of numerous smaller
calibration activities
• Calibration results from the most recent activity are marked in bold
• Calibration results originating from a previous activity are marked
in light gray color
• If a calibration result is generated in connection with a
measurement, there is a link to the related result
• If a 2-point calibration is initiated, all results are recent activities
and therefore marked in bold
Automated corrective actions
• A comprehensive evaluation ensures that specifications are met
and that calibration and QC are correct
• If rejected, the calibration/QC is automatically repeated. All actions
are logged and traceable.
Levey-Jennings plot for quality control
• Overview of day-to-day QC variations
• All the measured parameters are checked at low, normal and high
concentration levels, covering the relevant physiological measuring
range
• QC values that are comparable even if you change a solution pack
• The gray bar marks the change of solution pack and QC lot
changes: