Syringe samples – preanalytical errors, continued
Sample collection and handling
5. Visually inspect sampler and expel all air bubbles
• Air bubbles can affect O
2
and CO
2
values
safePICO syringes: Expel air bubbles through the vented safeTIPCAP
• Cap syringe to avoid contact with room air
6. Manual mixing or mixing using the integrated
mixing device
Stress that mixing makes the sample more homogeneous:
• Mixing reduces sedimentation
• Hemolysis may cause K
+
↑, Na
+
and Ca
2+
• When mixing manually, invert the syringe according to local guidelines
• Discuss how to mix a sedimented sample, follow local guidelines.
• There is no real consensus as to how to invert a syringe –
an easy way to mix manually is by rolling the syringe between
your palms
• Stress that the integrated sample mixer is for safePICO
syringes only
4. Correct volume of blood
• Too much blood = increased risk of clotting
• Too little blood = possible heparin binding effect
(may cause lower Ca
2+
value)
• Flush solutions used in A-lines must be removed from the
sample site to avoid dilution of blood samples
• It is recommended to withdraw a volume equal to three to
six times the “dead space” of the catheter system (CLSI)
Be aware of possible mixtures of venous and arterial blood: If
a self-filling syringe does not fill quickly, it may be because the
needle punctured the back of the arterial wall and continued
into a vein (with a lower pressure). In that case, a new sample
should be taken.
7. Analyze quickly
• If not analyzed quickly, the sample will continue to metabolize, causing the following:
pO
2
since O
2
will still be consumed
pCO
2
since CO
2
will still be produced
pH due to change in CO
2
cCa
2+
↑ the change in pH will influence the binding of Ca
2+
to protein
cGlu since glucose will be metabolized
cLac due to metabolism
6 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals