23
Replacements, continued
Consumables Default tests or Recommended replacement
activities per day interval after installation
Solution pack 10 Maximum 30 days or when the
number of activities is zero
Sensor cassette 10 Maximum 30 days or when the
number of tests is zero
Inlet gasket holder 10 12 months
Inlet connector gasket 10 12 months
Can a sensor cassette be used again?
A sensor cassette removed from one analyzer can be used on the same or on another
ABL90 FLEX analyzer if the following six conditions are met:
• The sensor cassette is kept right side up after its removal. This prevents damage
to the sensors.
• The sensor cassette is installed within 2 hours of its removal
• The sensor cassette is installed before its Scheduled-to-replace date
• The sensor cassette is installed before its Expiration date
• The sensor cassette has some remaining tests
• The sensor cassette was not removed from an analyzer during a long-term shutdown
procedure
This data can be seen in the sensor cassette status screen.
Can a Solution Pack be used again?
A Solution Pack removed from one analyzer can be used on another if these 3 conditions
are met:
• the Solution Pack is installed before its Scheduled to replace: date
• the Solution Pack is installed before its Expiration date:
• the Solution Pack has some remaining activities
This data can be seen in the Solution Pack Status screen.
For further instructions, please refer to the latest version of “ABL90 FLEX Instructions for use”
(under “Replacement and maintenance”).
23 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019