RADIOMETER FLEX ABL90 - Analyzer Parameters; Measured Parameter Ranges

RADIOMETER FLEX ABL90
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameters
31
Measured parameters
Ranges of indication and reportable ranges can be found in the ABL90 FLEX Instructions for use.
Be sure to always consult the latest version.
The Range of indication for a parameter is the range within which the analyser is physically capable
of measuring.
Reportable range is the range of results from a testing system or method over which the analytical
performance is claimed. The default reportable ranges stated in the ranges over which Radiometer
has validated the analytical performance and obtained clearance with the US FDA.
*) The values are for analyser with activated 'Out of range' suppression
31 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019