Changing consumables – solution pack
• Show how to enter Analyzer Status on the analyzer screen and find
> Consumables > Replace > Replace Solutions.
• Lift the inlet handle to the capillary position and wait until the solution
pack is released
• Let the trainee remove the solution pack and dispose of it as
biohazardous waste according to local guidelines
• Take a new solution pack and demonstrate how get it out of the box
without pulling or lifting in the pump tube
• Lift the new solution pack out of its box as shown and place it on an
even surface
• Now let the trainee try to get the solution pack out of the box
• Let the trainee pull the red pin out of the new solution pack
• Show how to place the palms of your hands over the edges of the
lid as shown
• Now let the trainee place their palms and tell them to press down
firmly and evenly with both hands until the tabs click into the two holes
• Stress that both taps must click into place for the solution pack to be
activated correctly
17
Changing consumables, continued
17 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019