EasyManuals Logo

RADIOMETER FLEX ABL90 User Manual

RADIOMETER FLEX ABL90
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Changing consumables – solution pack
• Show how to enter Analyzer Status on the analyzer screen and find
> Consumables > Replace > Replace Solutions.
• Lift the inlet handle to the capillary position and wait until the solution
pack is released
• Let the trainee remove the solution pack and dispose of it as
biohazardous waste according to local guidelines
• Take a new solution pack and demonstrate how get it out of the box
without pulling or lifting in the pump tube
• Lift the new solution pack out of its box as shown and place it on an
even surface
• Now let the trainee try to get the solution pack out of the box
• Let the trainee pull the red pin out of the new solution pack
• Show how to place the palms of your hands over the edges of the
lid as shown
• Now let the trainee place their palms and tell them to press down
firmly and evenly with both hands until the tabs click into the two holes
• Stress that both taps must click into place for the solution pack to be
activated correctly
17
Changing consumables, continued
17 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals