4. Introducing the sample
Present the following steps on how to introduce the sample.
Then the trainee should repeat the steps as a practical exercise.
Inform the trainee that the analyzer will guide with on-screen video,
sound and instructions during the process.
Key words:
LIFT – PUSH – HOLD – REMOVE – CLOSE
LIFT
• Lift the inlet to the “Capillary” position until the green light appears.
PUSH
• Push the capillary tube onto the inlet, applying pressure to depress
the inlet gasket
HOLD
• Hold the capillary tube in place while the sample is automatically
aspirated
REMOVE
• Gently remove the capillary tube from the inlet when prompted
CLOSE
• Close the inlet
Capillary samples – how to measure
a patient sample, continued
13 of 36EXT-VEND-MAN-0023-1 Owner:Julie Monnox Approved:27 Mar 2019

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER FLEX ABL90 and is the answer not in the manual?

RADIOMETER FLEX ABL90 Specifications

General IconGeneral
Analyzer TypeBlood Gas Analyzer
Analysis Time35 seconds
User InterfaceTouch screen
Power Supply100-240 V AC, 50/60 Hz
Parameters MeasuredpH, pCO2, pO2
Sample TypeWhole blood
Sample Volume65 µL
CalibrationAutomatic
ConnectivityUSB, Ethernet
Operating Temperature15-30°C
Operating Humidity20-85% relative humidity, non-condensing

Related product manuals