EasyManuals Logo

RADWAG AS 310.R2 User Manual

RADWAG AS 310.R2
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
- 60 -
Automatic adjustment process takes place in three different areas:
Adjustment on plugging the balance to mains verified balances.
Adjustment triggered by temperature change inside the balance. The
balance is equipped with very precise system for monitoring
temperature. At each adjustment process, the temperature is saved
in the system. The next adjustment is automatically initiated if
temperature changes more than 3°C from the last saved
temperature.
Adjustment triggered by elapsing time. It is possible to declare time
intervals which are criteria for balance adjustment. Accessible
settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours
since last carried out adjustment procedure,
CAUTION
Declaring time intervals for balance adjustment is available only in
non-verified balances. In case of verified balances the time interval
between the following adjustments is set for 3 hours.
13.2. Menu adjustment settings
P1 CALIBRATION
P1.1 INT.CALIB. | [internal adjustment]
P1.2 EXT.CALIB. | [external adjustment]
P1.3 USER CALIB. | [user adjustment]
P1.4 CALIB. TEST | [adjustment test]
P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH [automatic
adjustment]
P1.6 AUTO CAL. C. | [time of automatic adjustment]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS 310.R2 and is the answer not in the manual?

RADWAG AS 310.R2 Specifications

General IconGeneral
BrandRADWAG
ModelAS 310.R2
CategoryScales
LanguageEnglish

Related product manuals