The total mass of all the pieces put on the weighing pan must not be greater
than the balance’s maximum weighing range.
The mass of a single piece cannot be lower than the 0,1 reading unit of the
balance. If this condition is not met, the balance displays the message: <Too
low sample mass>.
While determining the number of pieces, wait for the stability pictogram ,
in order to confirm the quantity.
Only after the stability pictogram is displayed can you confirm the declared
quantity by pressing
button. Otherwise, the balance will not accept the
measurement.
17.2.3. Setting the reference mass by entering its mass
Procedure:
Press one of the F buttons, to which the function <ENTER SAMPLE>
is assigned, and the editing window <ENTER SAMPLE> is displayed
Use the arrow buttons to insert the known a single sample mass value.
Confirm the entered mass by pressing button,