- 130 -
The parameter enables a user to use the drying mode settings as in case of the
moisture analyzers MAC and MAX series. Thus, on purchasing a moisture
analyzer MA 3Y series it is not necessary to determine the drying parameters for
the products to be dried, but simply use the selected and proved drying modes.
Procedure:
Enter submenu < Others>
Select parameter <Compatibility with MAC/MAX> which opens a window
with drop down menu,
Select one of available options:
o YES parameter enabled
o NO parameter disabled
Selecting the parameter value causes immediate returning to the menu
window.
23.11. Sensors sensitivity
The parameter is adjustable between 0 9, and it determines the
distance from which the IR proximity sensors should react.
As standard, the sensitivity is set within the range of 5 to 7.
Procedure:
Enter submenu < Others>
Select parameter <Sensor sensitivity> which opens an editing window,
Select one of available options. The selection causes immediate returning
to the menu window.
23.12. Proximity sensors delay
Parameter for setting delay of proximity sensors, it is expressed in [ms].
By default the value is set to <0>.
When using option of motion recognition the user must remember about setting <proximity
sensors delay> parameter to value <500ms>
Procedure:
a) Go to < Other> submenu.
b) Select <Proximity sensors delay> parameterand wait for a dialog
window with an on-screen keyboard to open.
c) Enter value <500>.
d) Confirm. The program returns to menu window.
23.13. Autotest
Option permanently disabled in the moisture analyzer MA 3Y series.