- 129 -
Caution:
Means of leveling the moisture analyzer is described in point 11.3 of this user
manual.
23.9. Drying temperature reducing
This parameter automatically determines reducing the drying temperature after
a time interval specified for a moisture analyzer with Max drying temperature of
250 °C.
Means of operation:
While drying temperature holds within a limit between 161 °C and 250 °C then
the drying temperature can be maintained proportionally to set temperature
value, i.e. about 15 hours for the temperature 161 °C and about 10 minutes for
the temperature 250 °C.
While drying a sample in a temperature set to 250 °C, this temperature can be
maintained for the time of 10 minutes, and afterwards the software
automatically reduces the temperature to 160 °C (drying process is continued
without break).
The time of reducing the temperature from 250 °C to 160 °C is about 10
minutes.
For the QUICK MODE the initial exceeding of the Max drying temperature is 30
%, but not higher than the maximum drying temperature of a moisture analyzer.
Procedure:
Enter submenu < Others>
Select parameter <Reduction of drying temperature> which opens a new
window with selectable values
Select one of the available options:
YES temperature reduction mode enabled
NO temperature reduction mode disabled
Selecting a value causes returning to the menu window.
Caution:
Remember that prolonged drying a sample is temperature above 160
°
C may
cause damaging the components of the moisture analyzer.
23.10. Compatibility with MAC/MAX series
This parameter enables reflecting operation of moisture analyzers MAC and MAX
series. By enabling the parameter (i.e. settings its value to ,YES>), the moisture
analyzer MA 3Y series starts functioning (in the field of drying temperatures and
control of the drying process) as the moisture analyzers MAC and MAX series (in
comparable temperature conditions).