PROCEDURE 1 - on taking the tarred load off moisture analyzer’s weighing
pan
Press ZERO key
The NET marker is deleted, and new zero point of the moisture analyzer is
determined
PROCEDURE 2 - when the tarred load in on moisture analyzer’s weighing
pan
Press ZERO key
The NET marker is deleted, and new zero point of the moisture analyzer is
determined
If tare value exceeds 2% of available maximal capacity, the display
indicates error message Err- (forbidden operation)
PROCEDURE 3 - when the tarred load in on moisture analyzer’s weighing
pan or on taking the tarred load off the weighing pan
Press programmable key <Deactivate tare>
The NET marker is deleted,
The display indicates tare value
Press <Restore tare> key to restore the last tare value
11.6. Additional parameters on weighing process
Changes in settings of functions related to weighing process can, to some
extend, influence moisture analyzer’s reactions.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts, Profile
3. Press one of available submenu options and select item for modification,
Menu <Settings> - additional options related to weighing mode or drying
mode
Menu <Keys> - defining quick access keys
Menu <Information> - supplementary data on weighing /drying process
displayed in the grey coloured workspace
Menu <Printouts> - selection of printout type
Menu <Profile> - selection of a profile, active during moisture analyzer
operation
Menu <SETTINGS> contains supplementary data on weighing process,
such as: