- 42 -
12. ADJUSTMENT
Moisture analyzers MA 3Y series feature adjustment (calibration) option, which
ensures their correct measurement accuracy. Menu <ADJUSTMENT> contains
functions controlling operation of moisture analyzer’s adjustment process,
including options:
12.1. External adjustment
External adjustment is carried out using an external adjustment weight, with
appropriate accuracy and mass relating to moisture analyzer’s type and
maximum capacity. The process takes place semi-automatically, and the
following process phases are indicated on terminal’s display.
Process course:
Enter submenu < Adjustment > and select option: “ External
adjustment”,
The display indicates the following message box:
If there is a load on moisture analyzer’s weighing pan, unload it,
Press key. The moisture analyzer determines start mass, which is
indicated by a message box: “Start mass determination” on the
display,
On determining the start mass, the moisture analyzer displays a message
box:
Acting according to displayed command, place an ordered weight/mass
standard on the weighing pan and press key,