EasyManuals Logo

RADWAG MA 3Y User Manual

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
- 36 -
SINGLE,
mass value stored in instrument’s memory on single pressing of TARE
key, the following pressing of TARE key determines new value of tare.
Selection of product or packaging with assigned tare value causes
automatic deleting the previously assigned tare value;
SUM OF ACTIVE,
totalizing applied tare values of a product or a packaging (resulting from
selecting a product or a packaging from corresponding database) with a
possibility of increasing determined tare value by manual tare introduced
manually using moisture analyzer’s numeric keyboard. On following
determining of tare value for a product or a packaging, the tare value
introduced manually is deactivated;
SUM OF ALL,
totalizing all introduced tare values;
AUTOTARE
Means of operation:
Each first stable measurement result is tarred. The NET pictogram
flashes on the display. The operator can determine net mass of weighed
load. On taking off the load from the weighing pan, and as the indication
returns to autozero zone, the software automatically deactivates Tare
value.
Automatic footer printout
Accessible options:
MODE
Nomanual footer printout,
Sum of measurements the condition for printing a footer is
exceeding mass value set in parameter <Threshold>
Number of measurements the condition for printing a footer is
carrying out a pre-defined number of measurements (a batch) set in
parameter <Threshold>
THRESHOLDsetting the value of threshold determining footer printout.
For option <Sum of measurements> the value is determined in
measuring unit [g], and for option <Number of measurements> the
value is not measurable, and determined by number of carried out
measurements.
Printout mode / value release
PRINT / VALUE RELEASE key, (manual operation of ENTER key)
Never printout deactivated
First stable the first stable measurement is printed
Each stable all stable measurements are set for printing
Each printout of all measurements (stable and unstable)
AUTOMATIC MODE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA 3Y and is the answer not in the manual?

RADWAG MA 3Y Specifications

General IconGeneral
BrandRADWAG
ModelMA 3Y
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals